Page 8EDIT/IN-SERVICE MENUEntering Edit Mode and Putting Meter In-ServiceFL_METER040720RWG15Tag:PV (Volume)Model Number:Analog Output Current:Percent
Page 9Exiting Edit Mode (Or putting the meter in-service) - To put the meter in-service, click the “Edit/In-Service” pull-down menu on the Main screen
Page 10SAVING/RESTORING CONFIGURATION MENUSave/Restore Conguration Meter SettingsRacine Federated HART Interface UtilityFL_METER040720RWG15Tag:PV (Vo
Page 11Save Conguration - This command will cause the meter to copy any new settings from RAM to EEPROM. At power up, settings are retrieved from EEP
Page 12Restore Factory Conguration - This command will cause the meter to load the factory settings into RAM and EEPROM. Use this command with cautio
Page 13CONFIGURE METER MENURacine Federated HART Interface UtilityFL_METER040720RWG15Tag:PV (Volume)Model Number:Analog Output Current:Percent of Rang
Page 14Analog Range Values (Re-scaling the meter)The Analog Range settings can be read and set from the “Analog Range Values” dialog box. To bring up
Page 15Primary Variable (PV)The primary variable can be read and set from the “Primary Variable” dialog box. To bring up this box, click the “Congure
Page 16HART InterfaceThere are many HART related elds that can be viewed and set using the “HART Interface” di-alog box. To bring up this box, click
Page 17Number of Preambles - This eld allows the user to change the number in the preambles at the beginning of HART messages transmitted by the mete
TABLE OF CONTENTSMinimum System Requirements ...1Installati
Page 18Process TemperatureThe Process Temperature settings are accessed through the “Process Temperature” dialog box. To bring up this box, click the
Page 19Integrated RTD - If the meter is supplied with an integrated RTD, it is typically the best choice for the “Process Temperature Source”. The int
Page 20Process PressureThe Process Pressure settings are accessed through the “Process Pressure” dialog box. To bring up this box, click the “Congure
Page 21Process Pressure Source - There are four different sources that can be selected for the pro-cess pressure: • An absolute pressure analog input
Page 22User Dened Pressure - This eld allows the user to enter a constant process pressure. However, this pressure is only used when “User Dened” i
Page 23The main purpose of this window is to allow the user to specify the source that the meter should use for density and viscosity information. Thi
Property Table Version Number - This eld displays the version number of the loaded prop-erty table.Process Density - This eld displays the process d
LCD Display ModeThe display mode can be accessed through the “LCD Display” dialog box. To bring up this box, click the “Congure Meter” pull-down menu
Alternating PV & Totalizer - This selection causes the LCD to alternate between the PV and totalizer. This selection is only valid when the primar
Output Current and Percent of Range - This selection causes the analog output current andthe percent of range values to be output during burst mode. C
MINIMUM SYSTEM REQUIREMENTSTo run the Racine HART Interface, the user must have the following:• Microsoft Windows 98, ME, NT, or XP. (the Racine HART
Password ProtectionPassword Verification:Password:WriteReadClosePassword Protection EnabledTo password protect the meter, click the “Password Protecti
The active alarm current setting is read from the meter and displayed when this dialog box is opened. To change the setting, click the desired selecti
For shedding frequencies below the selected cut-off, the output value will be set to 0 ow. This feature is helpful at low ows where noise in the env
Totalizer SettingsRoll Over Value:Multiplier:Custom Unit:WriteReadCloseTotalizer Enabled10000000.001The value of the totalizer is stored in the meter’
Totalizer Enabled - The totalizer is running when this box is checked. The totalizer can alsobe started and stopped by checking and un-checking this b
TOOLS MENURacine Federated HART Interface UtilityFL_METER040720RWG15Tag:PV (Volume)Model Number:Analog Output Current:Percent of Range:Process Tempera
Power Up - The settings of the totalizer are stored in non-volatile memory. Therefore, if the totalizer is running (totalizing) when the meter losses
Page 35The “Property Table” dialog box is a tool used to load property tables into the meter. This dialog box can also be used to export a property ta
Page 36DiagnosticsThe diagnostics screen is typically only necessary when the meter needs to be serviced. A Racine customer service technician can ref
Save AsFile name:Save as type:Save in:SaveCancelMy ComputerMy Network P...My DocumentsDesktopHistoryMy DocsText.txt??PrintCancelPrintPage RangeAllSele
THE MAIN SCREENThe Racine HART Interface main screen displays basic meter measurements and status indicators. The main screen also provides menus that
Conguration ReportThe conguration report is a convenient tool to display all the customer congurable settings of the meter. The report generated by
CalibrationThe calibration selection is only necessary when the meter needs to be serviced. The Racine customer service technician will utilize or dir
Page 40The “Digital Pots” selection will allow the technician to change hysteresis data.CloseWriteReadDigital Pot SettingsUnfiltered Audio Hyst:Pot Va
Page 41APPENDIX A - WIRING DIAGRAMSUse foil shielded cable in all installations. Ground the foil shield to the chassis ground terminals provided on th
Analog Pressure Gauge WiringIf an analog pressure gauge is used it must be wired as shown below to avoid inadvertent cur-rent paths.A single supply ca
APPENDIX B - TROUBLE SHOOTING GUIDEProblem ResolutionLow loop voltageThe meter requires 15 volts DC at the power terminals. Measure this voltage at th
A division of Racine Federated Inc.Limited Warranty and Disclaimerracine, div. of Racine Federated Inc. warrants to the end purchaser, for a period of
8635 Washington AvenueRacine, WI 53406-3738 USA262-639-6770888-572-2463e-mail [email protected] 45
Page 3• Fixed Current Mode - This indicator is lit when the analog output current is set to a xed value. The analog output current is not dependent
Page 4Pull Down Menus• Congure Meter - This menu gives the user access to the basic meter settings.Racine Federated HART Interface UtilityFL_METER04
Page 5• Tools - This menu provides tools to the user such as analog output trimming and totalizer. Racine Federated HART Interface UtilityFL_METER040
Page 6• Save/Restore Conguration - This menu allows the user to save and recall setting from non-volatile memory (EEPROM).Racine Federated HART Inte
Page 7• About - This selection opens a dialog box that presents the Racine HART Interface soft-ware version and also provides contact information for